Por favor, use este identificador para citar o enlazar este ítem:
https://repositorio.ufba.br/handle/ri/7201
metadata.dc.type: | Artigo de Periódico |
Título : | Multivariate optimization and validation studies in on-line pre-concentration system for lead determination in drinking water and saline waste from oil refinery |
Otros títulos : | Microchemical Journal |
Autor : | Santos, Walter L. dos Santos, Carla M. M. dos Costa, Jorge L. O. Andrade, Heloysa Martins Carvalho Ferreira, Sergio Luis Costa |
metadata.dc.creator: | Santos, Walter L. dos Santos, Carla M. M. dos Costa, Jorge L. O. Andrade, Heloysa Martins Carvalho Ferreira, Sergio Luis Costa |
Resumen : | The present paper proposes an on-line pre-concentration procedure for lead determination in drinking water and saline waste from oil refinery by flame atomic absorption spectrometry (FAAS). It is based on the sorption of lead (II) ions in a minicolumn of polyurethane foam loaded with 4-(2-pyridylazo)-resorcinol (PAR) reagent. The optimization step was performed using Doehlert matrix involving the variables: sampling flow rate (SR), buffer concentration (BC), pH and eluent concentration (EC). The validation process was performed considering the parameters: linearity and other characteristics of the calibration curve, analytical features of on-line system, precision, robustness, effect of other ions in the pre-concentration system and accuracy. Using the established experimental conditions, the procedure allows lead determination with detection limit (3δ/S) of 0.4 μg l−1, quantification limit (10δ/S) of 1.4 μg l−1, and a precision, calculated as relative standard deviation (RSD) of 5.7 (n=8) and 2.1% (n=8) for lead concentration of 5 and 50 μg l−1, respectively. The pre-concentration factor (PF) considering the ratio among the slopes of the analytical curves with and without pre-concentration is 51. The achieved recovery for lead determination in presence of several cations demonstrated that this procedure could be applied for analysis of water samples. The accuracy was confirmed by analysis of the standard reference material NIST 1640 Trace elements in natural water. The sorption process was characterized by the Langmuir isotherm. The method was applied for lead determination in drinking water collected in Salvador City, Brazil and in saline effluent samples from oil refinery. The lead content for 16 samples of drinking water analyzed varied from 0.77 to 6.98 μg l−1. |
Palabras clave : | On-line pre-concentration system Lead determination Oil refinery |
URI : | http://www.repositorio.ufba.br/ri/handle/ri/7201 |
Fecha de publicación : | 2004 |
Aparece en las colecciones: | Artigo Publicado em Periódico (Química) |
Ficheros en este ítem:
Fichero | Descripción | Tamaño | Formato | |
---|---|---|---|---|
1-s2.0-S0026265X04000505-main.pdf Restricted Access | 335,94 kB | Adobe PDF | Visualizar/Abrir Request a copy |
Los ítems de DSpace están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.