Use este identificador para citar ou linkar para este item: https://repositorio.ufba.br/handle/ri/4896
Tipo: Artigo de Periódico
Título: Construção de uma cela de fluxo para medidas por espectrofotometria em fase sólida
Título(s) alternativo(s): Química Nova
Autor(es): Reis, Boaventura F.
Rocha, Fábio Rodrigo Piovezani
Teixeira, Leonardo Sena Gomes
Costa, Antonio Celso Spinola
Korn, Mauro
Autor(es): Reis, Boaventura F.
Rocha, Fábio Rodrigo Piovezani
Teixeira, Leonardo Sena Gomes
Costa, Antonio Celso Spinola
Korn, Mauro
Abstract: Construction of a flow cell for measurements by solid-phase spectrophotometry. A simple and low cost flow cell is proposed for measurements by solid-phase spectrophotometry employing a conventional spectrophotometer. The flow cell geometry allows the employment of a large amount of the solid support without causing both excessive attenuation of the radiation beam and increasing of the back-pressure. The adaptation of the flow cell in the optical path of the spectrophotometer in order to increase the precision is discussed. The flow cell characteristics were demonstrated by measurements of Co(II), employing 1-(2-tiazolylazo)-2-naphthol (TAN) immobilized on C18 bonded silica as solid support. The apparent molar absorptivity and coefficient of variation were estimated as 1.86 x 105 L mol-1 cm-1 and 1.4 % (n=15). A sample throughput of 40 determinations per hour and a detection limit of 15 mg L-1 (99.7 % confidence level) were achieved.
Palavras-chave: Flow cell
Solid-phase spectrophotometry
Flow analysis
URI: http://www.repositorio.ufba.br/ri/handle/ri/4896
Data do documento: 2000
Aparece nas coleções:Artigo Publicado em Periódico (Química)