Title: Optimization of fluorine determination via the molecular absorption of gallium mono-fluoride in a graphite furnace using a high-resolution continuum source spectrometer
Other Titles: Spectrochimica Acta Part B: Atomic Spectroscopy
Authors: Gleisner, Heike
Welz, Bernhard
Einax, Jürgen W.
metadata.dc.creator: Gleisner, Heike
Welz, Bernhard
Einax, Jürgen W.
Abstract: The determination of fluorine using the molecular absorption of gallium mono-fluoride (GaF) at the 211.248 nm rotational line has been optimized using a commercially available high-resolution continuum source atomic absorption spectrometer with a transversely heated graphite tube furnace. The electron excitation spectrum of GaF was generated by adding 500 μg Ga per injection into the graphite tube as molecule forming reagent. Best results were obtained by applying Zr as a permanent modifier and a mixed Pd/Zr modifier, thermally pretreated before each sample injection together with the Ga reagent at 1100 °C. The use of sodium acetate and Ru(III) nitrosyl nitrate as additional modifiers injected together with the sample further improved the performance. This way a maximum pyrolysis temperature of 550 °C could be used, and the optimum molecule forming temperature was 1550 °C. Several drinking water samples, a mineral water sample, and two certified reference materials were analyzed using the standard calibration technique; the absence of potential matrix effects was verified by measuring different dilutions and spiking with known fluorine mass. The results were in good agreement with the certified values and those measured by ion selective electrode; the recovery rate for the spiking experiments was between 97% and 106%. The results show that there was no matrix influence for that kind of samples containing relatively high concentrations of Ca, Mg and chloride, which are known to cause interference in GaF molecule absorption. The limit of detection and the characteristic mass of the method were 5.2 and 7.4 pg F, respectively, and were both about a factor of two better than recently published values.
Keywords: High-resolution continuum source AAS (HR-CS AAS)
Molecular absorption spectrometry (MAS)
Fluorine determination
Gallium mono-fluoride
Graphite furnace
URI: http://www.repositorio.ufba.br/ri/handle/ri/12541
Issue Date: 2010
Appears in Collections:Artigo Publicado em Periódico (Química)

Files in This Item:
File Description SizeFormat 
1-s2.0-S0584854710002375-main.pdf
  Restricted Access
397,68 kBAdobe PDFView/Open Request a copy


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.